SET FOR SEROLOGICAL DIAGNOSIS OF BOVINE PARAINFLUENZA-3 IN HEMAGGLUTINATION INHIBITION REACTION (HI)
Main characteristics of the kit
- Trade Name
- SET FOR SEROLOGICAL DIAGNOSIS OF BOVINE PARAINFLUENZA-3 IN HEMAGGLUTINATION INHIBITION REACTION (HI)
- Contents of the kit
-
- Hemagglutinating antigen to parainfluenza virus type 3
- Specific serum against parainfluenza virus type 3
- Control serum
- Indications for use
- Detection of antibodies to bovine parainfluenza-3 virus in hemagglutination inhibition reaction
Description
The active ingredient in the diagnostic kit is a concentrated twin-ether parainfluenza-3 virus antigen.
Release form
The diagnostic kit is available in liquid form.
Composition
The diagnostic kit includes the following components:
- Hemagglutinating antigen for PG-3 virus – 1 vial in a volume of 1.5 cm³
- Specific serum for PG-3 virus – 1 vial of 1.0 cm³
- Control serum – 1 bottle with a volume of 1.0 cm³
Indications
The kit is designed to detect antibodies to the parainfluenza-3 virus of cattle in the hemagglutination inhibition reaction (HI).
Method of application
Blood serum samples from sick and recovered animals are used as the samples to be analyzed, at the beginning of the disease and 3-4 weeks later.
At least 15-20 serum samples are sent for research.
The sera are stored under conditions that prevent bacterial growth at a temperature of 2 to 8°C for no more than 15 days.
For longer storage (up to 2 months), boric acid can be added as a preservative (2-3 crystals per 1 cm³ of serum). After dissolution, diagnostic hyperimmune serum can be stored at 2-8°C in a test tube with a rubber stopper for up to 10 days.
Serum can be stored at minus 20°C for a year, and it should be thawed at room temperature before testing.
Paired sera should be examined simultaneously (regardless of the results of the first examination), and the first serum should be stored frozen before examination.
Serum samples with pronounced hemolysis, bacterial growth, or those that have been stored for a long time without freezing or have been repeatedly frozen are not allowed.
All samples to be examined must be identified (labeled).
Marking of components
Labels are attached to the vials containing the following information: the name of the manufacturer, its address and trademark, the name of the component, the volume in the vial, the batch number, the date of manufacture, the working dilution of the component, the expiration date, and the designation of the Standard.
According to the package, the bottles are packed in cardboard boxes according to GOST 33781 or in plastic (polystyrene) containers with dividing partitions that ensure their stability and integrity.
Each box contains an instruction on how to use the diagnostic kit. Each package with a set of vials is labeled with the following information: the name of the manufacturer, its address and trademark, phone number, name of the components, number of vials in the box, serial number, date of manufacture (month, year), expiration date (month, year), number of tests, storage conditions, and the designation of the Standard.
Preparation of components and solutions for assay
Antigen titer verification. 0.05 cm³ of physiological saline with a pH of 7.2-7.4 is added to all wells of the microplate. 0.05 cm³ of antigen is added to the first wells of each row and titrated for a twofold dilution from 1:2 to 4096. Then, 0.05 cm³ of a 1% suspension of guinea pig red blood cells is added to all wells. Blood is taken from the guinea pig's heart and placed in a flask containing a 1.0% suspension of sodium citrate in a physiological solution with a pH of 7.2-7.4. The red blood cells are washed in a physiological solution with a pH of 7.2-7.4 and a 1.0% suspension is prepared.
4AE antigen test.
The antigen titer is divided by 4 (for example: if the titer is 512:4=128, then the antigen must be diluted by 128 times). To do this, take 1.0 cm³ of antigen and 127.0 cm³ of physiological solution with a pH of 7.2-7.4. Shake well.
To test 4AE, a physiological solution with a pH of 7.2-7.4 of 0.05 cm³ is added to all wells. Next, a pre-diluted antigen is added to the first wells of each row, titrated, and a 1.0% suspension of guinea pig red blood cells is added to all wells. The mixture is shaken and left at room temperature for 2 hours.
For diagnostic purposes, the test serum is diluted 1:4 with distilled water, transferred to a sterile test tube, and heated for 30 minutes at 56°C.
Performing Analysis
The hemagglutination inhibition reaction is carried out using a micro-method in polystyrene plates with U-shaped wells for single-use immunological reactions.
0.05 cm³ of physiological solution with a pH of 7.2-7.4 is added to the wells of the microplate. The first wells of each row are filled with 0.05 cm³ of the serum being tested. Two rows of wells are required for each serum being tested. The subsequent transfers result in a twofold dilution of the serum, ranging from 1:2 to 4096. Then, antigen (4AE) is added to each well, the panel is shaken, and left for 60 minutes at 37°C.
After 60 minutes, 0.05 cm³ of 1.0% guinea pig red blood cell suspension is added to all wells, shaken, and left at room temperature for 2-3 hours.
The reaction is accompanied by the following controls:
a) control for the absence of spontaneous agglutination: physiological solution + guinea pig erythrocytes.
b) control: saline solution + control serum + guinea pig red blood cells.
c) control: physiological solution + specific serum + guinea pig red blood cells.
The diluted antigen should be stored at 2-8°C for 24 hours, and the undiluted antigen should be stored in a freezer (-18-20°C) for 12 months.
Registration of results
The reaction is counted after 3 hours. The highest dilution of the serum that gives a clear inhibition of agglutination (button) is taken as the titer of antibodies in the IH. A positive serum is one that gives a titer of 1:16 or higher in the IH.
a) control wells: saline solution + guinea pig red blood cells – the reaction should be negative.
b) in the control wells, the titer of saline solution + control serum + guinea pig erythrocytes should not exceed 1:8.
in the control wells, the titer of saline solution + specific serum + guinea pig red blood cells should be at least 1:128.
An increase in titers in paired sera by a factor of four or more, as well as an increase in the number of seropositive sera, indicates the presence of infection.
Procedure for filing a complaint
In the event that the diagnostic set does not meet the requirements specified in this manual, or if it is ineffective, the use of this series of the product should be discontinued. The manufacturer is responsible for monitoring the diagnostic set that is subject to complaints.
Special Precautions
When conducting research, it is necessary to:
- work in boxed rooms for microbiological research using personal protective equipment (protective clothing and disposable rubber gloves);
- avoid contact with skin and mucous membranes, and immediately rinse with plenty of water if contact occurs;
- do not pipette with your mouth;
- in the event of spills of samples and working solutions on working surfaces, it is necessary to carry out disinfection treatment using disinfectants whose effectiveness has been confirmed in relation to the pathogen used in the work;
- tools and equipment (before and after use) should be treated with 70% ethyl alcohol or disinfectants that have been proven to be effective against the pathogen in question;
- the remaining test serum samples and polystyrene plates are disinfected by immersing them in a 3% chloramine solution for 5 hours or by autoclaving at 0.5 atm for 30 minutes;
- remnants of hemagglutinating antigen do not require special disposal conditions and can be poured into the sewer.
Manufacturer
Agrovet LLC, 109472, Moscow, Tashkentskaya Street, 34, Building 5